Industrial weights

Calibrating M1, M2 and M3 weights

M-class weights are used to check the scales in production, warehouse and shop: bench scales, floor scales, pallet scales and weighbridges. We calibrate knob weights, block weights and ECO weights in classes M1, M2 and M3. Where necessary we also clean, repaint and adjust them, all in-house.

Cast iron block weights in class M1

Which designs do we calibrate?

M-class weights come in many shapes and materials. We calibrate regardless of brand:

Knob weights

Stainless steel or chrome-plated knob weights from 1 g to 20 kg, usually with an adjustment cavity under the knob.

Block weights

Rectangular cast iron or stainless steel weights from 5 kg to 50 kg and larger, with a grip or lifting eye and a sealed adjustment cavity.

ECO weights

Stackable block weights with a compact shape. They are practical for testing floor and pallet scales.

Heavy weights

Mass pieces and block weights from 50 kg to 2,000 kg are calibrated using our own lifting and transport equipment.

Read more

Permissible error in M1, M2 and M3

The M classes have wider tolerances than the F classes, but here too every gram counts. A 20 kg block weight in class M1 may deviate by no more than 1 gram.

Nominal valueF2M1M2M3
1 g0.30 mg1.0 mg3.0 mg10 mg
100 g1.6 mg5 mg16 mg50 mg
1 kg16 mg50 mg160 mg500 mg
5 kg80 mg250 mg800 mg2.5 g
10 kg160 mg500 mg1.6 g5 g
20 kg300 mg1 g3 g10 g
50 kg800 mg2.5 g8 g25 g
100 kg1.6 g5 g16 g50 g
500 kg8 g25 g80 g250 g
1,000 kg16 g50 g160 g500 g

Maximum permissible error (±) according to OIML R111-1.

Which M class for which scale?

The weight must be more accurate than the scale you test with it. The common rule from OIML R76 is that the error of the test weight is at most one third of the error the scale is permitted at that load. In practice this means:

  • M1 for most class III scales: retail, bench, floor and pallet scales and many weighbridges;
  • M2 for coarser class III scales and class IIII scales;
  • M3 for coarse industrial applications and as a loading weight, for example in proof load tests.

If in doubt, choose the finer class. You can also use an M1 weight later if you purchase a more accurate scale. For the calibration of scales according to EURAMET cg-18, the measurement uncertainty of the weight also determines the final result. Read more under scale calibration.

Cleaning, repainting and adjusting

With us, a calibration of M-class weights is a complete overhaul of the test weight: inspection, cleaning, derusting and repainting where necessary, adjustment and recalibration. This way a cast iron weight lasts for years to come.

In practice, M-class weights are used heavily: they stand on floors, are moved with forklifts and come into contact with dust, moisture and oil. At intake we therefore first assess the condition of each weight.

  • Cleaning: we remove dirt, oil and loose rust before measuring. Otherwise the dirt is calibrated along with the weight.
  • Repainting: cast iron weights with a damaged paint coat are treated and given a new protective coat. This prevents corrosion and the resulting loss of mass.
  • Adjusting: block and knob weights have an adjustment cavity. We add or remove adjustment material until the weight is within tolerance, close the cavity again and measure again.

Because all these steps take place in our own laboratory and our own workshop, weights do not have to go back and forth between different parties. That saves time and prevents damage in transit.

Weighing scheme and certificate

We compare M-class weights with reference weights of a higher class, using the OIML R111 weighing schemes. Our own LIMS software records every measurement, corrects for air density and calculates the conventional mass with its measurement uncertainty.

For each weight, the certificate states the measured value, the deviation, the measurement uncertainty and the assessment in its class. If a weight was adjusted, the values before and after adjustment are stated. The certificate is also available in your customer portal, together with the date of the next calibration.

Frequently asked questions

Can rusty cast iron weights still be calibrated?

Usually yes. We remove loose rust, treat the weight and apply a new protective coat. We then adjust and calibrate it. If the weight is so badly affected that it no longer remains stable, we advise replacement.

What does adjusting a block weight cost?

That depends on the deviation and the condition of the weight. We let you know in advance if adjustment is necessary, so there are no surprises. Please contact us for a quote.

Which class do I need for my floor scale?

For most floor and pallet scales in class III, M1 is the right choice. For coarse scales or class IIII, M2 is sufficient.

Can an M1 weight be calibrated in a higher class?

A weight can only be assessed in a class if it also meets the requirements for material, surface and shape of that class. A cast iron M1 weight therefore does not become an F2 weight, even if its deviation is small.

Do you also collect large quantities of weights?

Yes. We collect weights with our own transport and equipment, even pallets full of block weights, throughout the Netherlands and Belgium.

Buy weights

New calibration weights, optionally with a certificate from our own laboratory.

Related services

From the knowledge centre

All articles →

Schedule a calibration or verification?

Request a quote or call us. We calibrate in our own laboratory in Geldermalsen or on site, throughout the Netherlands and Belgium.

View basket →