Precision weights

Calibrating F1 and F2 weights

F1 and F2 weights are the reference for precision balances, laboratory scales and the finer industrial scales. In these classes, tenths of a milligram matter. We calibrate them in our climate-controlled room, using reference weights of a higher class and the OIML R111 weighing schemes.

Weight set with precision weights in a case

What are F1 and F2 weights used for?

Class F1 is the usual class for checking and calibrating precision balances and for the finer class III scales. Class F2 suits precision balances with a coarser readability and good bench and retail scales. Both classes are also used as references for calibrating weights in the M classes.

F-class weights are usually made of stainless steel, polished or fine-turned, as knob weights, compact cylindrical weights or sheet weights below 1 gram. They are supplied individually or as a weight set in a case or box with tweezers and gloves.

Permissible error in F1 and F2

The maximum permissible error per nominal value according to OIML R111-1, compared with the neighbouring classes:

Nominal valueE2F1F2M1
1 mg0.006 mg0.020 mg0.06 mg0.20 mg
10 mg0.008 mg0.025 mg0.08 mg0.25 mg
100 mg0.016 mg0.05 mg0.16 mg0.5 mg
1 g0.030 mg0.10 mg0.30 mg1.0 mg
10 g0.060 mg0.20 mg0.60 mg2.0 mg
100 g0.16 mg0.5 mg1.6 mg5 mg
500 g0.8 mg2.5 mg8 mg25 mg
1 kg1.6 mg5 mg16 mg50 mg
5 kg8 mg25 mg80 mg250 mg
10 kg16 mg50 mg160 mg500 mg
20 kg30 mg100 mg300 mg1 g

Maximum permissible error (±) according to OIML R111-1. The expanded measurement uncertainty of the calibration may be at most one third of this value.

What does an F-class calibration require?

The finer the class, the more the conditions matter. For a 100 gram F1 weight the result may deviate by only half a milligram. The measurement uncertainty must then stay below 0.17 mg. That is only achievable if everything is right:

  • Temperature: weight, reference and mass comparator must be at the same temperature. A difference causes air flow around the weight and therefore a measurement error. Our calibration room is controlled at 20 ±0.5 °C and the weights acclimatise first.
  • Air density: a stainless steel weight and a reference with a slightly different density experience different air buoyancy. Our R111 software uses the measured air pressure, temperature and humidity and corrects for this.
  • Weighing scheme: we weigh according to ABBA or ABA, with several cycles per weight. This eliminates drift and random variation. The software flags outliers and checks repeatability.
  • Handling: we never touch F-class weights with bare hands. We use tweezers, gloves and clean surfaces so that no grease or moisture gets onto the weight.

Calibrating weight sets

A weight set from 1 mg to 1 kg easily consists of twenty or more weights. Duplicate nominal values, such as two 2 g or two 200 g weights, are marked, for example with a dot or an asterisk. We follow that marking so that each weight can be found on the certificate.

The case or box is part of the set. We check that all compartments are complete, that the tweezers still close properly and that the lining is clean. We report damaged sheet weights or deformed wire weights below 1 gram immediately, as these can often no longer be reliably adjusted. If a weight is missing, we can supply it separately and calibrate it in the same round.

Adjusting F-class weights

Many F1 weights are made from a single piece, without an adjustment cavity. If such a weight is slightly too heavy, it can sometimes be carefully corrected by fine polishing. If it is too light, no material can be added and adjustment is not possible. An F2 knob weight with an adjustment cavity under the knob can be adjusted in both directions.

Adjustment is not always the best choice. A weight that falls just outside F1 often still comfortably meets F2. We discuss with you whether downgrading to a lower class, adjusting or replacing is the most sensible option. That depends on what you use the weight for.

Storage between two calibrations

Most of the deviations we see in F-class weights arise between two calibrations. Always store weights in their case or box, in a dry room with a stable temperature. Handle them only with tweezers or gloves. Place them on a clean surface and never directly on a cold or damp workbench. Read more tips in our article on storing calibration weights.

Frequently asked questions

How often should F1 weights be calibrated?

Your quality system determines the interval. Annual calibration is customary for F1 weights. If successive results remain stable, you can extend the interval on a documented basis.

Can an F1 weight that falls outside tolerance still be used?

Often yes, but in a lower class. A weight that falls just outside F1 usually still meets F2. You can also continue to use it with the measured value from the certificate, if your procedure allows this.

Do you also calibrate weights below 1 gram?

Yes. We calibrate sheet and wire weights from 1 mg in classes F1 and F2. Because of their size they are susceptible to damage; we will tell you if a weight is no longer reliable.

Are the weights in my set listed separately on the certificate?

Yes. Each weight gets its own line, with the measured conventional mass, the deviation, the measurement uncertainty and the assessment.

Can I send in my F-class weights myself?

Yes, provided they are well packed in the original case or box. You can also have us collect them, throughout the Netherlands and Belgium.

Buy weights

New calibration weights, optionally with a certificate from our own laboratory.

Related services

From the knowledge centre

All articles →

Schedule a calibration or verification?

Request a quote or call us. We calibrate in our own laboratory in Geldermalsen or on site, throughout the Netherlands and Belgium.

View basket →